Technical specification:

1 . Air temperature:

Air temperature display range: 0℃-60℃;

Air temperature control range: 20.0℃-37.0℃, 37.1℃-39.0℃

Air temperature accuracy: ≤±0.3℃;

Air temperature deviation alarm: ±3.0℃.

2 . Skin temperature:

Skin temperature display range: 0℃-60℃;

Skin temperature control range: 20.0℃-37.0℃, 37.1℃-38.0℃;

Skin temperature accuracy :≤±0.3℃;

Skin temperature deviation alarm: ±1.0℃.

3. Humidity

Humidity display range: 0%--99%RH;

Humidity control range: 30%--95%;

Humidity control accuracy: (10% - 90% at 20 - 40℃)±5%RH.

4. Oxygen concentration (Optional)

Oxygen concentration display range: 18%-99%

Oxygen concentration display accuracy: 2%

Product feature:

● LCD display screen, easily set, clear, and intuitive;

● Air temperature display and servo control;

● Skin temperature display and servo control, double skin temperature sensor;

● Humidity display and servo control;

● Display time and heater power;

● 24-hour Parametric trend diagram display;

● Audible & Visual Alarms: power failure, sensor failure, higher & lower temp, air & skin temperature override, fan failure, low water level, system failure;

● >37℃ Preset Mode: power failure reset memory, alarm silence, self-test function;

● Hood with single slope, big front door, 6 access ports, 4 tube grommet;

● 2 Storage cabinets, Protective guard rail, IV poles;

● RS-232 Output Connector (used for network system);

● Optional oxygen concentration measure system;