• Infant Radiant Warmer VNT-2000B
  • Infant Radiant Warmer VNT-2000B

Infant Radiant Warmer VNT-2000B

The neonatal radiation table works based on the principle of radiant heat. It transfers heat energy to the infant through the infrared radiation device so that the infant can maintain a stable body temperature. Healthcare professionals can adjust the power and distance of the radiation according to the baby's needs to ensure that the baby's body temperature is appropriate.
  • Description

  • Trust points

TECHNICAL SPECIFICATIONS:    

  • Temperature Control Range: 25℃-37℃,
  • Temperature override mode to 38℃
  • Temperature sensor accuracy: ≤0.3℃
  • Mattress surface temperature uniformity: ≤2.0℃
  • Temperature rise time: ≤45mins
  • Heat output indicates: 0-100% in 10%
  • Radiant box horizon angle adjust range: 0-90°adjustable in step-less
  • Bed tilt adjustment: continues adjustable in ±10°
  • Mattress Tray Size: 1700px*1350px
  • Mattress tray Max. load: 10kg
  • Transfusion pole Max. load: 2kg
  • Shelf Max. load:2kg

ALARM

  • Sensor failure alarm
  • Over temperature alarm
  • Deviation alarm: ±1℃
  • Power failure alarm
  • APGAR timealarm with audible when run between 50 seconds to 1 minute, 4 minutes 50 seconds to 5 minutes,
  • 9 minutes 50 seconds to 10 minutes.

ENVIRONMENT TEMPERATURE

  • (It is a suggestion that please do not use outside of the allowed environment.)
  • Operate temperature range18℃~30℃
  • Store temperature range-40℃~+55℃
  • ENVIRONMENT HUMIDITY
  • Operate humidity range30%RH~75%RH
  • Store humidity range≤93%RH

ATMOSPHERIC PRESSURE

  • Transport and Store atmospheric pressure range500hPa~1060hPa
  • Operate atmospheric pressure range700hPa~1060hPa

STREAM VELOCITY

  • Ambit environment stream velocity≤0.3m/s

Features:

●Temperature Control Mode: pre-heating control, servo control & manual control;

● Instant Parameter Display: set temp, skin temp, heat power rate;

● Built-in APGAR Timer and >37℃ preset function

● Audible/Visual Alarms: temp deviation, sensor failure, power failure, temp override;

● Tilting Bed with Rotary Radiant Head;

● X-ray Cassette Tray under infant bed, examination light of high illuminance;

● Use an anti-blast quartz infrared tube as a radiant heating source;

● With two LED phototherapy units;

● With two 7L oxygen cylinders, with pressure meter and flow meter;

● With Suction, with neonate resuscitation.

  • Infant Radiant Warmer VNT-2000B